GHM SIM-1 SensorSimulator introduced to SA
Hãng sản xuất: GHM Group
Product no .: SIM-1
general information:
The GHM Sensor Simulator outputs various voltage and current signals. The GHM SensorSimulator can therefore also simulate sensors such as Pt100, thermocouples and strain gauges optimally due to the additional back-up measurement of the connected voltage amplifiers. Optionally a frequency output is available. It can be used to calibrate and verify displays, transducers or complete measurement chains. In addition, voltages and currents can be measured with the device.
specifications:
connections: | 7-pole Binder socket for signal input and output Micro-USB for supply voltage and charging |
display: | graphic LCD, monochrome, (180 x 128 pixels), adjustable background illumination |
operation : | keypad |
languages: | German, English |
dimensions: | 160 x 86 x 37 mm (H x W x D) |
weight: | 250 g (accu incl.) |
supply voltage: | 5V DC (Micro-USB) |
accu: | lithium ion |
ambient temperature: | 0 ° C to + 50 ° C |
simulation function | |
voltage: | simulation range: ± 10 V accuracy: ± 1% |
signal current: | simulation range : ± 25 mA accuracy : ± 1% |
strain gage bridge: | simulation range : 0, 0.5, 1, 2, 4, 5, 10, 25, 50 mV / V accuracy : ± 1% supply: 2.5 V, 5 V, 10 V |
thermocouple, type K (others opt.): | simulation range : -100 ° C to +1,000 ° C (steps: -100 ° C to + 100 ° C: 10 ° C + 100 ° C to + 500 ° C: 25 ° C + 500 ° C to + 1,000 ° C : 50 ° C) accuracy : simulated compensation: ± 1%; internal temperature measurement: ± 3K |
thermocouple, type J (others opt.): | simulation range : -100 ° C to +1,000 ° C (steps: -100 ° C to + 100 ° C: 10 ° C + 100 ° C to + 500 ° C: 25 ° C + 500 ° C to + 1,000 ° C : 50 ° C) accuracy : simulated compensation: ± 1%; internal temperature measurement: ± 3K |
thermocouple, type N (others opt.): | simulation range : -100 ° C to +1.250 ° C (steps: -100 ° C to + 100 ° C: 10 ° C + 100 ° C to + 500 ° C: 25 ° C + 500 ° C to + 1,250 ° C : 50 ° C) accuracy : simulated compensation: ± 1%; internal temperature measurement: ± 3K |
thermocouple, type S(others opt.): | simulation range : -50 ° C to +1,600 ° C (steps: -50 ° C to + 100 ° C: 10 ° C + 100 ° C to + 500 ° C: 25 ° C + 500 ° C to + 1,600 ° C : 50 ° C) accuracy : simulated compensation: ± 1%; internal temperature measurement: ± 3K |
Pt100: | simulation range : -100 ° C to +850 ° C (steps: -100 ° C to + 100 ° C: 10 ° C + 100 ° C to + 500 ° C: 25 ° C + 500 ° C to + 850 ° C : 50 ° C) accuracy : ± 1% |
frequency (option F): |
simulation range : 1 Hz to 500 kHz (steps: 1 Hz to 10 Hz: 1 Hz 10 Hz to 100 Hz: 10 Hz 100 Hz to 1 kHz: 100 Hz 1 kHz to 10 kHz: 1 kHz 10 kHz to 100 kHz 10 kHz 100 kHz to 500 kHz: 100 kHz) level (adjustable): ± 10 V accuracy : ± 1% |
measuring function | |
voltage: | measuring range: ± 30 V accuracy : ± 0.5% |
Electricity: | measuring range : ± 30 mA accuracy: ± 0.5% |
For further details prt data sheet: